Model |
RC-ZNBXAⅡ |
Pumping Mechanism |
Curvilinear peristaltic |
IV Set |
Compatible with IV sets of any standard |
Flow Rate |
1-1200ml/h (in 0.1ml/h increments) |
Purge, Bolus |
Purge when pump stops, bolus when pump starts, rate at 500, 800, 1200ml/h |
Accuracy |
±3% |
VTBI |
1-20000ml (in 0.1ml increments) |
Infusion Mode |
ml/h, drop/min, time-based, body weight, nutrition |
KVO Rate |
1-5ml/h (in 0.1ml/h increments) |
Alarms |
Occlusion, air-in-line, door open, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm |
Additional Features |
Real-time infused volume, automatic power switching, mute key, purge, bolus, system memory, history log |
Occlusion Sensitivity |
High, medium, low |
History Log |
50000 events |
Air-in-line Detection |
Ultrasonic detector |
Wireless management |
Optional |
Drop Sensor |
Optional |
Vehicle Power (Ambulance) |
12±1.2V |
Power Supply, AC |
110/230V (optional), 50-60Hz, 20VA |
Battery |
9.6±1.6V, rechargeable |
Battery Life |
5 hours at 25ml/h |
Working Temperature |
10-30℃ |
Relative Humidity |
30-75% |
Atmospheric Pressure |
860-1060hpa |
Size |
130*145*228mm |
Weight |
2.5kg |
Safety Classification |
Class Ⅰ, type CF |