Local alarm system

<?=$ecms_gr[title]?>
Model: RC-LAS01

Local alarm system

Fill in the form to get more product information

Product Introduction

Medical area gas pressure alarm is suitable for monitoring the gas supply pressure of hospital floor, nurse station, operating room and other areas, real-time digital display of the monitored gas pressure, alarm points can be set according to the requirements of each parameter, and the over-limit alarm can be realized to ensure the normal gas supply pressure of the hospital. This type of gas pressure alarm in medical area is equipped with RS-485 communication interface, which can realize the networking between the main alarm box and the system host through the network. 
• Host power supply voltage: DC9V±5%
• Power adapter power supply voltage: AC100V ~ 240VAC 
• Power consumption: 2W(1 gas), 3W(2 gas), 4W(3 gas), 5W(4 gas), 6W(5 gas), 7W(6 gas), 8W(7 gas) 
• Default voltage measurement range: -0.1MPa ~ 1.0MPa
• Pressure monitoring number: 1 ~ 7 
• Default pressure unit: MPa(positive pressure), kPa(negative pressure) 
• Support positive pressure detection of negative pressure sensor 
• Support RS-485 remote communication 
• Support switching quantity output of each air pressure state (instead of electric contact pressure gauge) 
• Support DC1 ~ 5V analog quantity output of each air pressure value (corresponding positive pressure: -0.1MPa ~ 1.0MPa, negative pressure: -0.1MPa ~ 0MPa) 
• Support switch output of the whole machine state
• Mute the infrared remote control within 5 meters
• Sensor fault detection
• Sound and light alarm
• Mute the buzzer
• All parameters can be debugable on site