Anesthesia machine

<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
Model: RC-AM888

Anaesthesia machine

Fill in the form to get more product information

Product Introduction

TECHNICAL SPECIFICATION
Ventilation mode
SIMVf/2,SIGH, SIPPV, DEMO, MANUAL
Ventilator parameter range
Flowmeter:02(0.05~10 L/min)
                 N20(0.05~10 L/min)
                 AIR (0.05~10 L/min)
Rapid oxygen supply: 35 L/min ~75 L/min
Tidal volume (Vt):0,20 mL~1500 mL
Frequence (Freq):4~99bpm
PEEP+0.5KPa--+2KPa
Pressure triggering sensitivity (PTr): -20 cmH²О ~ 0cmH0 (Based on PEEP)
Flow trigger sensitivity (FTr): 0.5 L/min ~ 30 L/min
Pressure control (PC): 5 cmH0 ~ 60 cmH2O

Alarm and protection
The AC power failure alarm
Internal battery backup low voltage alarm
No tidal volume
High oxygen concentration alarm
Low oxygen concentration alarm
High Airway pressure alarm
Low Airway pressure alarm
High Minute Volume alarm
Low Minute Volume alarm
Continuous Pressure alarm
Suffocation warning
The maximum limited pressure
Fan error
Oxygen deficit
Monitoring parameter
Frequence (Freq): 0 /min ~ 100 /min
Tidal volume (Vt):0 mL~2000 mL
MV: 0 L/min ~ 100 L/min
Oxygen concentration:15%~100 %