TECHNICAL PARAMETERS
PART OF ANESTHESIA MACHINE
*Main Body
High-strength engineering plastic rack,light, beautiful andcorrosion resistant
*Scope of Application:Adult, Child
*Gas Source02:0.28-0.6MPa,N20:0.28-0.6MPa
*Flow Meter:02:0.05-1.0L/min.1.1-10L/min N20:0.05-1.0Lmin,1.1~10L/min
*02-N20 Linkage and N20 Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0,16+0.02Kpa, the flow of nitrous oxide would be cut off.
*Flow Rate of Rapid Oxygen Supply 25~75L/min
*Low Oxygen Pressure AlarmThere will be sound alarm when the oxygen pressure < 0.2MPa
*Vaporizer
It has the function of automatic compensation based onpressure,temperature, and fow rate. The regulation range ofevaporator concentration is 0~5 vol%.
Among Halothane,Enflurane, lsoflurane and Sevoflurane,twocan be chosen for application as required by the customer.
*Respiratory Circuit
Working mode: All-close, Semi-close, Semi-open APL: ≤12.5 kPa
*Respiratory BellowsBellows for adults and childrenTidal volume range: 0~1500 ml
*Light
Equipped with LED illumination, easy for observation
*ACGO Port
With ACGO port to connect with T circuit or BAlN circuit foropen-operation.
*AGSS PortAGSS exhaust gas collection system can be installed,
PART OF VENTILATOR
*Display Mode High-definition 10.1*TFT color LCD touch screen display, easy!and convenient operation
*Ventilation Mode
IPPV, SIPPV, SIMV, PEEP, SIGH, MANUAL
*Ventilation Function PEEPSIGH,IRV
*Ventilation Parameters
Tidal volume:50~1500 ml
Rate:2-99 bpm
SIMV rate:2-20 bpm
I:E:2:1~1:8
Inspiratory trigger pressure: -10~20 cmHzO
PEEP:2-20 cmH,O
Pressure range:5~60 cmHzO
SlGH:1.5 times the inspiratory time/ 100 times breath
*Parameters For Ventilation Monitoring
Tidal Volume, Ventilation Volume, IPPV Rate, SlMV Rate, l:E,PeakPressure of Airway,Pressure-Time Waveform, Flow Rate-TimeWaveform, PEEP, Inspiratory Trigger Pressure (Average Pressure,Fi02 Oxygen Concentration, Battery Level)