Defibrillator

<?=$ecms_gr[title]?>
Model: RC-AED7000(plus)

Automatic External Defibrillator

Fill in the form to get more product information

Product Introduction

  • Product Description

    RC-AED7000(plus) Product Description of Defibrillator machine is a portable model, which can be equipped at home, in public places, or in the hospitals.

    It's easy and convenient to use when give first aid to patient.

    Meanwhile, it has the function of automatic analyzing patient's ECG data, and then takes corresponding defibrillation energy level according to patient's current situation, which has greatly improved success rate, and maximum reduced damage to patient's heart.

  • Features

    1. Color Screen size: 3.5"

    2. Automatic self-testing

    3. Biphasic energy output

    4. CPR metronome

    5. CPR Coaching (Optional)

    6. Adult and Pediatric defibrillation

    7. Extensive voice and visual prompts for the operator

  • Technical Specifications

    1. Color Screen size: 3.5"

    2. Automatic self-testing

    3. Biphasic energy output

    4. CPR metronome

    5. CPR Coaching (Optional)

    6. Adult and Pediatric defibrillation

    7. Extensive voice and visual prompts for the operator

  • Technical Specifications
    Category Specification
    Qutput Biphasic Truncated exponential
    Energy Sequence Adult: 150J, 150J, 200J
    Pediatric: 50J, 50J, 85J(ajustable)
    Charge Time 8sec. to 150J 4sec. to 50J/100J
    10sec. to 200J(New battery)
    Analysis Time 9 sec.
    The maximum time from the initiation of rhythm analy-sis to readiness for discharge with a new battery. Less than 20 seconds
    The maximum time from the initiation of rhythm analy-sis to readiness for discharge after 6 shocks. Less than 25 seconds
    The maximum time fore initially switching power on to readiness for discharge. Less than 30 seconds
    Audible Prompts 19 audible prompts
    Visual Prompts LED prompts; Automatic Self-test
    Controls Six Buttons: ON/OFF, Shock and four software configurable buttons
    Output Energy Accuracy ±15% into any impedance from 25 to 175Ω
    Output disabled when the patient impedance is outside limits 20Ω to 200Ω