1. CPAP ventilation modes keeps children spontaneous breath and reduces children breath work.
2. Continues to provide CPAP mode of constant flow and limited pressure for newborn baby.
3. Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child which will be highly improve the safety and efficiency of oxygen therapy.
4. Equipped with build-in air blender, ensuring O2 concentration adjustment to be stable.
5. Reliable CPAP control valve and pressure monitoring system improve CPAP adjustment precision.
6. Optional parts: humidifier with heating devices.
7. High cost-effective and reliable quantity.
8. Easy operation, compact appearance and design let user could operate more easily.
1) Parameter adjustment | |
CPAP adjusting range | 0 cmH2O ~ 10 cmH2O |
Continuous flow adjusting range | 1.5 L/min ~ 15 L/min |
Oxygen Concentration | 21%-100% |
2)Parameter monitoring range | |
CPAP/airway pressure | |
Monitoring range | 0 cmH2O ~ 60 cmH2O |
Accuracy | ±(2 % full scale + 4 % actual value) |
Continuous flow | |
Monitoring range | 1.5 L/min ~ 15 L/min |
Accuracy | ±4 % |
3) Gas supply | |
Gas source | O2(For medical use), Air(For medical use) |
Rated working pressure | 350 kPa |
Output pressure of regulator | 350 kPa±50 kPa (300 kPa~400 kPa) |
Ambient temperature | 5 ℃~ 40 ℃ |
Relative humidity | ≤80 % |
Atmospheric pressure | 960 hPa~1040 hPa |
Storage and transportation environment | |
Ambient temperature | -10 ℃~+40 ℃ |
Relative humidity | ≤90 % |
Atmospheric pressure | 860 hPa~1060 hPa |